Lucas c – Physio-Control LUCAS Manuel d'utilisation

Page 26

Advertising
background image

26

100359-50 B, ©J

OLIFE

AB 2006

LUCAS

TM

I

NSTRUCTIONS

FOR

U

SE

G

EBRUIKSAANWIJZING

I

NSTRUCTIONS

D

UTILISATION

2.2 U

SING

LUCAS C

HEST

C

OMPRESSION

S

YSTEM

WARNING – IMPROPER USE

Improper use of the device can cause serious injury

to the patient and ineffective chest compressions.

Jolife AB strongly recommends that the equipment

only be used by personnel who have undertaken

training in use and handling of LUCAS Chest

Compression System.

Other personnel may assist those with the

necessary training in using LUCAS.

WARNING – DEVICE STATUS

If your LUCAS unit does not appear as in the

illustration in Section 1.4, or if the device does not

function in the same way as in training, or according

to specification:

Discontinue use and continue with manual chest

compressions.

WARNING – MALFUNCTION

If there is air leakage, interruption in compressions

or insufficient compressions, if the frequency is

wrong, or there is jarring, or there is something else

that cannot be classified as normal running:

Disconnect the air hose immediately, turn the

ON/OFF

knob to

Adjust

and remove the device.

Start manual chest compressions.

WARNING – SUCTION CUP SLIDING ON GEL

If there is gel on the patient’s chest (e.g. from

ultrasound examination), the position of the suction

cup may change during use.

Remove any gel left before applying the suction cup.