Lucas c – Physio-Control LUCAS Manuel d'utilisation

Page 42

Advertising
background image

42

100359-50 B, ©J

OLIFE

AB 2006

LUCAS

TM

I

NSTRUCTIONS

FOR

U

SE

G

EBRUIKSAANWIJZING

I

NSTRUCTIONS

D

UTILISATION

3.5 O

PERATING

LUCAS C

HEST

C

OMPRESSION

S

YSTEM

When the position of the suction cup is correctly
adjusted according to Section 3.4, LUCAS can be
switched on:

1.

Turn the

ON/OFF

knob to

Active

.

LUCAS

will now give chest compressions.

WARNING – PATIENT INJURY OR DEATH

Do not leave the patient or device unattended while

LUCAS is active.

2.

Check that the device is working as it should
regarding frequency and compression.

3.

When you want to stop chest compressions,
turn the

ON/OFF

knob to the

Lock

position,

without changing the respective positions of
the patient and device.

Advertising