Physio-Control LUCAS Manuel d'utilisation

Page 56

Advertising
background image

56

100359-50 B, ©J

OLIFE

AB 2006

LUCAS

TM

I

NSTRUCTIONS

FOR

U

SE

G

EBRUIKSAANWIJZING

I

NSTRUCTIONS

D

UTILISATION

4.1.3 D

ISINFECTION

WARNING

– HAZARDS DURING

DISINFECTION

Always wear appropriate protective clothing during

disinfection of LUCAS. Follow the handling
instructions from the manufacturer of the

disinfectant.

The normal cleaning procedure is sufficient after
“normal” soiling. If LUCAS is blood-stained or if an
infectious patient has been treated, the device should
also be disinfected using 45% isopropyl alcohol with
added detergent.

4.1.4 D

ISINFECTION

PROCEDURE

After cleaning LUCAS, disinfect it using the following
procedure:

1.

Wipe all outer surfaces of the device using a
cloth wetted with the disinfectant solution.

2.

Allow LUCAS to dry before packing into the bag.

Advertising