Physio-Control LUCAS Manuel d'utilisation

Page 50

Advertising
background image

50

100359-50 B, ©J

OLIFE

AB 2006

LUCAS

TM

I

NSTRUCTIONS

FOR

U

SE

G

EBRUIKSAANWIJZING

I

NSTRUCTIONS

D

UTILISATION

1(

3.7.5 M

OVING

PATIENT

ON

TILTED

STRETCHER

In situations where conditions do not allow the
chest to remain horizontal:

1.

Turn the

ON/OFF

knob to

Lock

to pause

compressions (Figure

1(

).

2.

As soon as the patient is in horizontal position,
check the position of the suction cup and
adjust if necessary (see Section 3.4).

3.

Turn the

ON/OFF

knob to

Active

.

3.7.6 A

MBULANCE

TRANSPORT

Make sure that the patient is fixed when using
LUCAS Chest Compression System during
ambulance transport. The ambulance cot should
have four-point restraints for the patient’s upper
body in addition to the two lower-body restraints.

3.8 C

HANGING

AIR

SOURCES

WARNING – INTERRUPTED COMPRESSIONS

Excessive changeover time if regulators have to be

mounted.

Always use reserve cylinders with pre-fitted

regulators.

To change air source:

1.

Set the

ON/OFF

knob to

Adjust

.

2.

Close the valve on the used air cylinder.

3.

Disconnect the air hose from the air source.

4.

Open the valve on the new air source.

5.

Attach the air hose to the new air source.

6.

Adjust the height of the suction cup on the
patient’s chest, as described previously on
page 40.

7.

Turn the

ON/OFF

knob to

Lock

.

8.

Turn the

ON/OFF

knob to

Active

to

continue compressions.