g

3.6 D

EFIBRILLATION

WARNING – INCORRECT ECG ANALYSIS

Compressions interfere with ECG analysis.

Switch off LUCAS before ECG analysis. Make the

interruption as short as possible.

WARNING – INADEQUATE COMPRESSION

Patient’s movements may change adjustment of

suction cup.

Verify position of suction cup after defibrillation,

re-adjust if necessary.

Note: Self-adhesive electrodes should be used during
defibrillation as these make it easier to work with
LUCAS. Position the defibrillator electrodes and
wires so they are not under the suction cup; if
electrodes are already on the patient, you may need
to apply new electrodes.

Defibrillation can be performed when LUCAS is
applied to the patient (Figure

g

).

1.

Apply electrodes and perform the defibrillation
according to manufacturer’s instruction for the
defibrillator.

2.

After defibrillation or analysis of the outcome
verify the position of the suction cup and
re-adjust if necessary.

Advertising