3.7.3 L

IFTING

THE

PATIENT

At least three people should lift the patient and
LUCAS – one should support the patient’s head and
there should be one person on each side.

Plan carefully before lifting the patient. Consider
which equipment needs to be moved and where to
place the stretcher.

Those at the side should lift with one hand beneath
the claw locks of the back plate. The other hand
should be used to lift the patient’s belt, trouser or
under the thigh.

Use appropriate lifting technique to avoid injuries
when lifting the patient (Figure

k

).

WARNING – CRUSHING HAZARD

Do not insert your fingers in the claw locks.

To lift a patient:

1.

Secure the patient’s arms with the straps on the
support legs.

2.

Turn the

ON/OFF

knob to

Lock

, to pause

compressions while lifting the patient, for
example when:
– lifting to a stretcher
– moving from one stretcher to another

3.

When the patient has been placed on the
stretcher (or other surface), verify correct
position of the suction cup on the patient.

4.

Turn the

ON/OFF

knob to

Active

.

Once the patient is placed on the stretcher, LUCAS
may be in active state during all

horizontal lifts and

movements.

3.7.4 M

OVING

HORIZONTALLY

When the situation absolutely demands that
the patient be moved while undergoing chest
compression, the patient’s chest should be
horizontal throughout the entire movement
(Figure

l

).

Advertising