Rockwell Automation 1440-DYN02-01RJ XM Dynamic Measurement Module Manuel d'utilisation
Page 138

138
Publication ICM-UM002F-EN-E - March 2013
Appendix B CIP Objects
6
Get/Set
FMAX
REAL
The maximum frequency
or order of the spectrum
data.
0...20000 Hz if
Domain = 0.
There are several predetermined
FMAX settings for which
spectrum data can be produced.
If you select an unsupported
value, then the next greater
supported FMAX value will be
used for the spectrum data.
4...40 Orders if
Domain = 1.The
Number of Lines value must
be evenly divisible by the FMAX
value or an Invalid Device
Configuration error will be
returned during the Device
Mode Object Start and Save
services.
7
Get/Set
Number of
Spectrum Lines
UDINT
Number of lines or bins in
the spectrum data.
100, 200, 400, or 800
8
Get/Set
Window Type
USINT
The window function to
be applied to the
waveform data prior to
computing the spectrum.
0 = Rectangular
1 = Hamming
2 = Hanning
3 = Flat Top
4 = Kaiser Bessel
9
Get/Set
Period
REAL
The period of the
waveform.
Seconds if
Domain = 0.
Cycles if
Domain = 1.
10
Get
Number of
Waveform
Points
UDINT
Number of points in the
waveform data.
256, 512, 1024, or 2048
11
Get
Overlap
USINT
The percent overlap
applied to the waveform
data sets used for
calculating the spectrum.
Only 0% supported.
Table B.38 Spectrum Waveform Measurement Object Instance Attributes
Attr ID
Access
Rule
Name
Data Type
Description
Semantics