Error Messages

Error Messages

Error Messages

Error Messages

Error Messages

DISPLAYED

MESSAGE

DESCRIPTION

LobAt

Batteries are too weak to display weight values.

UnCAL

Invalid checksum of non volatile memory; unit must be set up for
program constants and calibrated. If error remains, it indicates a
hardware problem. Checksum is verified each time the

I

10 is

turned on.

ErrLo

Underload condition. Dead load weight is less than dead load
during calibration.

ErrHi

Weight is beyond 105% of full scale capacity.

PCErr

Parts counting error.

See Chapter 6.

Err0

Invalid memory checksum; indicates hardware error. Checksum is
verified each time the

I

10 is turned on.

Err2

Indicates overrange of A/D converter. Usually caused by over
weight condition beyond“ErrHi”.

Err3

RS232 Communication error - Invalid command sent.

Err5

Internal value is too large to display. Check Program Constants
menu or Custom Units Conversion Factor.

Err7

No weight units are enabled.

See Chapter 4.

tohi

Calibration/Linearization Error. Weight may have been added
during calibration procedure when no load should be on load
receiver.

toLo

Calibration/Linearization Error. Incorrect weight placed on load
receiver during calibration procedure.