To accept the displayed setting and pro-
ceed to the next parameter (Calibration
Point), press
.
To exit the Program Constants menu
and store your settings, repeatedly press
until SEtUP is displayed, then
press
again to return to the weigh-
ing mode.
Calibration Point
Calibration Point
Calibration Point
Calibration Point
Calibration Point
The calibration point is the weight value
that will be used to calibrate the
I
10. It
can be any value from 10% to 100% of
the Full Scale Point.
☞
If Linearization is enabled (inter-
nal Switch 1is ON), the calibration
point must be at least 75% Full
Scale Point.
To enter the calibration point:
1.
Access the “CALPt” parameter if it
is not already displayed.
2.
The display will show the current
calibration point with the first digit(s)
flashing. For first time setup, the
calibration point will be the same as
the full scale point entered.
3.
Press
to change the value
of the first digit.
4.
When the desired value is displayed,
press
to move to the next
digit.
Flashing
3
3
3
3
3
Setting Program Constants