Ohaus I10 Indicator Manuel d'utilisation

Page 47

Advertising
background image

47

Calibration

5

5

5

5

5

Calibration

Calibration

Calibration

Calibration

Calibration

5.

When “CAL 0” is displayed, press

.

“buSY” will be displayed briefly fol-
lowed by “CAL F” if 2 point calibra-
tion is being performed, or “CAL L”
if 3 point linearization is being per-
formed.

6.

For 2 point calibration, place a weight

equal to the Calibration Point on the

load receiver, then press

and

proceed to step 9.

For 3 point linearization, place a

weight equal to 1/2 the Calibration

Point on the load receiver, then

press

and continue with step

7.

7.

“buSY” will be displayed briefly, then
“CAL F” will be displayed.

8.

Place a weight equal to the Calibra-

tion Point on the load receiver, then

press

.

9.

“buSY” will be displayed briefly, then
“good”, then “SEtUP”.

When “SEtUP” is displayed, the

I

10 is

calibrated.

Press

to return to the weighing

mode.

or