Troubleshooting chart, Troubleshooting – Ohaus I10 Indicator Manuel d'utilisation

Page 63

Advertising
background image

63

Troubleshooting

Troubleshooting

Troubleshooting

Troubleshooting

Troubleshooting

PROBABLE CAUSE(S)

Power cord not plugged
in or properly connected.

Batteries dead or not
properly connected.

Load on load receiver ex-
ceeds allowable zero %
entered in “ZERO” pa-
rameter of Program Con-
stants menu

(see Chap-

ter 4).

Load receiver motion or
disturbances exceed
center of zero criteria.

Desired unit not set to
ON in Unit Selection
menu.

Display is negative.

REMEDY

Check power cord con-
nections.

Make sure adapter con-
nector is plugged all the
way into the indicator.

Check battery connector.
Replace alkaline batter-
ies or recharge nicad
batteries.

Remove load on load re-
ceiver to less than en-
tered zero %.

Change allowable zero
% in “ZERO” parameter
of Program Constants
menu

(see Chapter 4).

Remove disturbances or
reduce motion.

Increase averaging level
in AL menu.

Enable desired unit (see

Chapter 5).

Only KG and LB are avail-
able when LFT is set to
ON.

Remove load, zero indi-
cator, replace load and
tare again.

SYMPTOM

Unit will not turn on.

Cannot zero indicator, or
will not rezero when
turned on.

Center of Zero display
indicator erratic or does
not appear with no load
on load receiver.

Cannot display weight in
desired weighing unit.

Unable to tare (front
panel or RS232 com-
mand).