Calibration, Calibration procedure – Ohaus I10 Indicator Manuel d'utilisation

Page 46

Advertising
background image

46

Calibration Procedure

Calibration Procedure

Calibration Procedure

Calibration Procedure

Calibration Procedure

Make sure there is no load on the plat-
form.

1.

Press and hold

to access the

Setup mode.

Release

when “SEtUP” is

displayed.

2.

Press and hold

until “CAL” is

displayed, then release it.

Dead Load Adjustment

Dead Load Adjustment

Dead Load Adjustment

Dead Load Adjustment

Dead Load Adjustment

3.

When

is released,

“dLOAd” will be displayed, then the
current dead load setting.

4.

If dead load is “good”, then “CAL 0”

Adjustment
Potentiometer

will be displayed. Proceed to step 5.

If dead load is displayed as a per-

centage, for example “dL3.5%”, it

must be adjusted to be between

0.5% and 3.0% as follows:

Using the screwdriver provided, turn

the dead load adjustment potenti-

ometer clockwise to increase

clockwise to increase

clockwise to increase

clockwise to increase

clockwise to increase or

counterclockwise to decrease

counterclockwise to decrease

counterclockwise to decrease

counterclockwise to decrease

counterclockwise to decrease the

displayed value.

Press

to enter the dead load.

If deadload reads “ErrLo”, turn

potentiometer clockwise

clockwise

clockwise

clockwise

clockwise until per-

cent sign (%) appears, then adjust

between 0.5% and 3.0%.

5

5

5

5

5

Calibration