Setup and calibration restrictions, Labeling, Sealing – Ohaus I10 Indicator Manuel d'utilisation

Page 59

Advertising
background image

59

Setup and Calibration

Setup and Calibration

Setup and Calibration

Setup and Calibration

Setup and Calibration

Restrictions

Restrictions

Restrictions

Restrictions

Restrictions

When LFT is selected in the program
constants menu, the following param-
eter settings are automatically restricted
by the software:

Auto-Zero Tracking may be set at 0.6,
1 or 3.

Only pounds and/or kilograms may be
used.

RS232 Auto Print may be set to “At”
(OFF) or “StAb” (output on stability)
only.

Display resolution may be set to a
maximum of 10,000 divisions.

Calibration is restricted to 2 Point Cali-
bration.

Verify that the

I

10 has been configured

according to these restrictions.

The

I

10 capacity and readability

must match that of the NTEP cer-
tified Class III or IIIL or Canadian
Legal Metrology Branch General
Class load receiver it will be used
with.

Labeling

Labeling

Labeling

Labeling

Labeling

Type the capacity and resolution on the
capacity label provided in the accessory
kit and affix it to the recess above the
display.

Label

Legal For Trade Sealing

8

8

8

8

8