Full scale point – Ohaus I10 Indicator Manuel d'utilisation

Page 29

Advertising
background image

29

Full Scale Point

Full Scale Point

Full Scale Point

Full Scale Point

Full Scale Point

The maximum capacity of the load re-
ceiver is entered for Full Scale Point
using the same weighing unit selected
for the Calibration Unit.

If Legal for Trade is selected, reso-
lution will be limited to 10,000 divi-
sions. If you attempt to enter a full
scale point that will yield a resolu-
tion greater than 1:10,000, the dis-
play will flash “LFT”, then “Err”
(error), and return to the Gradua-
tions parameter.

To enter the Full Scale Point:

1.

Access the “FS” parameter if it is not
already displayed.

2.

The display will show the current full
scale point with the first digit(s) flash-
ing.

3.

To change the flashing digit(s), re-

peatedly press

until the de-

sired digit is displayed.

To shift to the next digit, press

.

To turn ALL digits ON and reset

them to zero, press

.

4.

Set all digits until the desired Full
Scale Point is displayed.

Divisions = FS Pt/Graduations

Flashing

Setting Program Constants

3

3

3

3

3

Advertising