Calibration

Calibration

Calibration

Calibration

Calibration

The

I

10 must be calibrated after

Program Constants have been ini-
tially set, and when any of the
Program Constants are changed

(see Chapter 3

- Program Con-

stants). Calibration is accessed
from the Setup mode.

Before beginning calibration you
will need to have the following
items on hand:

a calibration weight equal to the
Calibration Point entered in the
Program Constants menu (see

Chapter 3)

If linearization will be performed,
a weight equal to 1/2 of the
Calibration Point will also be
required

the screwdriver which was pro-
vided with the

I

10 (not required

for factory calibrated bench
scales; -C0 models).

Only 2 point calibration is
available for Legal For Trade
installations.

The rear cover of the

I

10 must be

removed before calibrating if the
SETUP switch is set to OFF. Refer
to

Chapter 1 i

f needed.

MODEL

CAL WT. (lbs)

LB30-C0

30

LB60-C0

60

PL100-C0

100

PL150-C0

150

WB10-C0

10

WB25-C0

25

WB50-C0

50

FACTORY CALIBRATED

FACTORY CALIBRATED

FACTORY CALIBRATED

FACTORY CALIBRATED

FACTORY CALIBRATED

BENCH SCALES

BENCH SCALES

BENCH SCALES

BENCH SCALES

BENCH SCALES

If your

I

10 is part of a factory calibrated

bench scale setup (-C0 model number
suffix), the weights listed below are
recommended to perform calibration.

5

5

5

5

5

Calibration