Gain jumper, Linearization switch – Ohaus I10 Indicator Manuel d'utilisation
Page 45

45
Gain Jumper
Gain Jumper
Gain Jumper
Gain Jumper
Gain Jumper
The gain setting for the
I
10 must match
the sensitivity of the load receiver being
used. Gain can be set for one of three
levels; 1 mV/V, 2 mV/V or 3 mV/V. Gain
is set at 2 mV/V when originally shipped.
Place the gain jumper in the appropriate
position on the rear panel for the desired
gain setting (see diagram).
Gain Jumper
Linearization Switch
Linearization Switch
Linearization Switch
Linearization Switch
Linearization Switch
If linearization is to be performed (3 point
calibration), Switch 1 on the rear panel
must be in the ON position. For 2 point
calibration, it should be OFF.
☞
3 point linearization is NOT avail-
able for Legal For Trade applica-
tions.
1mV/V
2mV/V
3mV/V
Switch shown in
ON postion.
Calibration
5
5
5
5
5