Gain jumper, Linearization switch – Ohaus I10 Indicator Manuel d'utilisation

Page 45

Advertising
background image

45

Gain Jumper

Gain Jumper

Gain Jumper

Gain Jumper

Gain Jumper

The gain setting for the

I

10 must match

the sensitivity of the load receiver being
used. Gain can be set for one of three
levels; 1 mV/V, 2 mV/V or 3 mV/V. Gain
is set at 2 mV/V when originally shipped.

Place the gain jumper in the appropriate
position on the rear panel for the desired
gain setting (see diagram).

Gain Jumper

Linearization Switch

Linearization Switch

Linearization Switch

Linearization Switch

Linearization Switch

If linearization is to be performed (3 point
calibration), Switch 1 on the rear panel
must be in the ON position. For 2 point
calibration, it should be OFF.

3 point linearization is NOT avail-
able for Legal For Trade applica-
tions.

1mV/V

2mV/V

3mV/V

Switch shown in
ON postion.

Calibration

5

5

5

5

5