Guidance and manufacturer’s declaration – electromagnetic immunity (continued)

The AutoSet CS2 is intended for use in the electromagnetic environment specified below. The customer or the
user of the AutoSet CS2 should assure that it is used in such an environment.

Immunity test

IEC60601-1-2
test level

Compliance
level

Electromagnetic environment – guidance

Portable and mobile RF communications
equipment should be used no closer to any part
of the AutoSet CS2, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.

Recommended separation distance

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80 MHz

3 Vrms

d = 1.17 √P

Radiated RF
IEC 61000-4-3

10 V/m
80 MHz to 2.5 GHz

10 V/m

d = 0.35 √P 80 MHz to 800 MHz

d = 0.70 √P 800 MHz to 2.5 GHz

where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,

a

should be less than the compliance level in each
frequency range.

b

Interference may occur in the vicinity of
equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the AutoSet CS2 is used
exceeds the applicable RF compliance level above, the AutoSet CS2 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the AutoSet CS2.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

eng.book Page 49 Friday, November 3, 2006 2:43 PM