ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation

Page 48

Advertising
background image

44

I

NSUFFICIENT

PRESSURE

SUPPORT

ALARM

Is triggered if for 3 breaths in a row the maximum mask pressure achieved at
inspiratory to expiratory transition is either <87.5% of the absolute mask pressure
targetted at that point, or < absolute mask pressure targetted minus 1 cmH

2

O,

whichever is the lower.

H

IGH

LEAK

ALARM

Leak exceeds 30 L/min for more than 20 seconds.

F

LOW

BLOCKED

ALARM

Flow is less than 6 L/min for more than 20 seconds.

IEC 601-1

CLASSIFICATIONS

Class II, Type CF
Class II medical electrical equipment provides protection against electric shock either

by double insulation or reinforced insulation and does not require a protective earth.

Table 3: Displayed values

Table 4: Maximum low impedance flow at stated pressures

Value

Range

Accuracy

Display
Resolution

Proximal cuff

Pressure

-3 to 40 cmH

2

O

0.5 cmH

2

O

0.1 cmH

2

O

Mass flow sensor

Leak

0–99 litres per
minute

+/- 12 litres per
minute

1 litre per breath

Respiratory rate

8–30 breaths per
minute

+/- 10%

1 breath per
minute

Tidal volume

0–4500 millilitres

+/- 10%

1 millilitre

Minute ventilation

0–40 litres per
minute

+/- 10%

0.1 litres per
minute

Target MV

0–40 litres per
minute

+/- 10%

0.1 litres per
minute

Pressure
(cmH

2

O)

Flow (L/min)

4.2

217

8.6

218

13.0

219

eng.book Page 44 Friday, November 3, 2006 2:43 PM

Advertising