ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation

Page 36

Advertising
background image

32

W

ILL

CLEAR

WHEN

:

the low pressure condition is fixed

treatment is stopped by pressing the

Start/Stop key.

P

OWER

FAIL

ALARM

T

RIGGERED

BY

:

power failure

machine is disconnected or switched off while delivering treatment.

I

F

A

POWER

FAIL

ALARM

SOUNDS

:

The flow generator stops delivering air pressure.

W

ILL

STOP

:

when the

Alarm Mute key is pressed, or

after 2 minutes, or

when power is restored.

!

CAUTION

Remove the mask from your face if the power fails.

H

IGH

PRESSURE

ALARM

T

RIGGERED

BY

:

the mask pressure exceeds the set trigger level (25 cmH

2

O) for more than 700

milliseconds.

I

F

A

H

IGH

PRESSURE

ALARM

SOUNDS

:

1. The treatment will stop.
2. Turn power off.
3. Check that the air tube and sensor tubes are connected properly.
4. Turn power back on.
5. Remove mask and perform Air Circuit Learn function.
6. Try using the flow generator one more time.
7. If the high pressure alarm activates repeatedly, discontinue use and return to

ResMed for servicing. If the alarm does not recur, then continue to use as normal.

N

OTE

The alarm system is designed to ignore coughing. However, if a cough is especially intense
and prolonged, it may trigger the high pressure alarm.

W

ILL

STOP

WHEN

:

the A

UTO

S

ET

CS2 is turned off.

H

IGH

LEAK

ALARM

T

RIGGERED

BY

:

high mask leak (greater than 30 L/min) for more than 20 seconds.

I

F

A

HIGH

LEAK

ALARM

SOUNDS

:

Adjust the mask to minimise leak.

eng.book Page 32 Friday, November 3, 2006 2:43 PM