Sing – ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation
Page 25

21
U
SING
THE
A
UTO
S
ET
CS2
E
NG
LI
SH
U
SING
THE
A
UTO
S
ET
CS2
!
WARNING
Before starting therapy with a new mask, select the correct mask type in the
Settings menu (page 26). When adding or removing a new component such as a
mask, humidifier or anti-bacterial filter, perform the Air Circuit Learn function
(page 26).
S
TARTING
T
REATMENT
When the system is fully assembled, press the power switch at the back of the
A
UTO
S
ET
CS2 to ON. The Welcome screen will be displayed while the flow
generator is warming up. Allow the unit to warm up for approximately 30 seconds
before placing the mask on your face.
Figure 12: Welcome screen
Once the A
UTO
S
ET
CS2 has warmed up, breathing into the mask should start therapy
immediately. If not, press the
Start/Stop key once to commence treatment.
The A
UTO
S
ET
CS2 restarts in the mode in which it stopped.
S
TOPPING
T
REATMENT
If S
MART
S
TOP
is enabled, treatment will stop automatically when you remove your
mask. If the S
MART
S
TOP
function has been disabled, press the
Start/Stop key.
S
TANDBY
M
ODE
Figure 13: Standby screen
Once you have stopped treatment, the A
UTO
S
ET
CS2 will be in standby mode. In
standby you can:
•
press the power switch at the back to OFF to turn off the flow generator
•
commence treatment again by breathing into the mask or pressing the
Start/
Stop key
•
enter any of the menus by pressing the
Left key.
When in standby mode, a gentle (non-therapeutic) flow of air will still be present.
N
OTE
If the SMART DATA - AUTO APPEAR option is set to ON, the screens available under the
Results menu will be displayed automatically for approximately 20 minutes after your
session is completed.
WELCOME
AUTOSET CS2
menu
eng.book Page 21 Friday, November 3, 2006 2:43 PM