CS2 unit is fitted with alarms to alert you to changes that will affect your

treatment. The most common reason for an alarm to sound is because the system has
not been properly assembled. Check that the air tubing and pressure sensor tube have

been properly attached to the flow generator and mask (and humidifier if used).
You can mute an alarm by pressing the

Alarm Mute key once. Unless the alarm

requires you to turn the power off, you can press any key again to remove the alarm
message from the LCD screen. If the problem is still present, the alarm will sound and
display again after one minute. The Alarm LED will remain lit for as long as the problem

is present.
All menus remain available during an alarm.
We recommend that you test the audible alarm once a week. To do this, hold down
the

Mute key while the A

UTO

S

ET

CS2 is starting up. If the alarm is functioning

correctly it will beep twice.

L

OW

MASK

PRESSURE

ALARM

T

RIGGERED

BY

:

pressure sensor line is blocked or disconnected

air pressure at the mask has fallen below a set level

mask is removed and SmartStop has been disabled.

I

F

A

LOW

PRESSURE

ALARM

SOUNDS

:

1. Check that the air tube and sensor tube are connected properly.
2. Turn the A

UTO

S

ET

CS2 off and on again at the power switch. If the alarm persists,

return the unit to ResMed for servicing.

LCD message

Alarm type

Action

LOW PRESSURE!
Check circuit

Low mask pressure alarm

Check tube connections.

LCD turns off

Power fail alarm

Turn off AutoSet CS2 at power switch.

HIGH PRESSURE!
Turn power off

High pressure alarm

Turn off AutoSet CS2 at power switch.

HIGH AVG PRESS!
Turn power off

High pressure alarm

Turn off AutoSet CS2 at power switch.

HIGH LEAK!
Check circuit

High leak alarm

Adjust the mask.

LOW PS!
Check circuit

Insufficient Pressure
Support alarm

Check tubing.
Run LEARN CIRCUIT.

FLOW BLOCKED!
Turn power off

Flow Blocked alarm

Turn off AutoSet CS2. Check air
circuit. Turn on AutoSet CS2.