The A
UTO
S
ET
CS2 unit is fitted with alarms to alert you to changes that will affect your
treatment. The most common reason for an alarm to sound is because the system has
not been properly assembled. Check that the air tubing and pressure sensor tube have
been properly attached to the flow generator and mask (and humidifier if used).
You can mute an alarm by pressing the
Alarm Mute key once. Unless the alarm
requires you to turn the power off, you can press any key again to remove the alarm
message from the LCD screen. If the problem is still present, the alarm will sound and
display again after one minute. The Alarm LED will remain lit for as long as the problem
is present.
All menus remain available during an alarm.
We recommend that you test the audible alarm once a week. To do this, hold down
the
Mute key while the A
UTO
S
ET
CS2 is starting up. If the alarm is functioning
correctly it will beep twice.
L
OW
MASK
PRESSURE
ALARM
T
RIGGERED
BY
:
•
pressure sensor line is blocked or disconnected
•
air pressure at the mask has fallen below a set level
•
mask is removed and SmartStop has been disabled.
I
F
A
LOW
PRESSURE
ALARM
SOUNDS
:
1. Check that the air tube and sensor tube are connected properly.
2. Turn the A
UTO
S
ET
CS2 off and on again at the power switch. If the alarm persists,
return the unit to ResMed for servicing.
LCD message
Alarm type
Action
LOW PRESSURE!
Check circuit
Low mask pressure alarm
Check tube connections.
LCD turns off
Power fail alarm
Turn off AutoSet CS2 at power switch.
HIGH PRESSURE!
Turn power off
High pressure alarm
Turn off AutoSet CS2 at power switch.
HIGH AVG PRESS!
Turn power off
High pressure alarm
Turn off AutoSet CS2 at power switch.
HIGH LEAK!
Check circuit
High leak alarm
Adjust the mask.
LOW PS!
Check circuit
Insufficient Pressure
Support alarm
Check tubing.
Run LEARN CIRCUIT.
FLOW BLOCKED!
Turn power off
Flow Blocked alarm
Turn off AutoSet CS2. Check air
circuit. Turn on AutoSet CS2.