Roubleshooting – ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation

Page 43

Advertising
background image

39

T

ROUBLESHOOTING

E

NG

LI

SH

T

ROUBLESHOOTING

If there is a problem, try the following suggestions. If the problem cannot be solved,

contact your equipment supplier or ResMed. Do not attempt to open the flow
generator.

Problem

Possible cause

Solution

Insufficient air delivered from
flow generator

Air filter is dirty.

Replace air filter.

Air tubing is kinked or
punctured.

Straighten or replace tubing.

Flow generator operating
erratically, ie, switching on and
off, pressure varying at random

Water may have entered the
air tubing.

Remove the water from the
air tubing.

Pressure rises inappropriately

You are talking, coughing,
voluntarily breath-holding (eg
while rolling over in bed), or
intentionally breathing in an
unusual manner.

Avoid talking with the mask
on, and breathe as normally
as possible.

Severe mask leak or mouth
leak (greater than 30 L/min).

Correct your leak condition.

Alarm stays on continuously

Blockage in air circuit.

Ensure there is no barrier to
the passage of air.

Internal malfunction.

Return the device to your
equipment supplier for
service.

No LCD display

Power not connected or
switch at back is not on.

Ensure the power cable is
connected, and that the
switch at the back of the unit
is in the “on” position.

FAULT message displayed on
LCD

Internal malfunction.

Return the device to your
equipment supplier for
service.

System Error 7 message

Device power-on has been
interrupted by breathing into
the mask during warm up.

Turn the flow generator off
and on again. Do not place
mask on face during warm up
(30 seconds). Select LEARN
CIRCUIT (page 26). If the
error is not cleared, contact
your equipment supplier for
service.

Internal malfunction.

Return the device to your
equipment supplier for
service.

eng.book Page 39 Friday, November 3, 2006 2:43 PM