ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation

Page 41

Advertising
background image

37

C

LEANING

AND

M

AINTENANCE

E

NG

LI

SH

!

WARNING

Do not wash the air filter once it has become soiled. The air filter is not washable
or reusable.

F

ITTING

THE

FILTER

Remove the filter cover at the back of the flow generator.

Remove and discard the old air filter.

Insert a new filter, the blue tinted side facing towards you.

N

OTE

Match the cut corner of the filter with the shape of the casing.

Replace the air filter cover. The blue tinted side of the filter should be facing out.

H

YPOALLERGENIC

AIR

FILTER

The ResMed hypo-allergenic filter is a standard filter to which an electrostatic filter has
been bonded to capture very small particles. The filter is intended for those users of

ResMed flow generators who will benefit from enhanced filtering of the air delivered
during treatment.
Inspect the air filter every month and check if it is blocked by dirt or contains holes.
The filter should be replaced every three months, or more often if the flow generator

is operating in a dusty environment.
Fit as described above.

S

ERVICING

This product (A

UTO

S

ET

CS2) should be inspected by an authorised ResMed Service

Centre 5 years from the date of manufacture. Prior to this, the device is intended to
provide safe and reliable operation provided that it is operated and maintained in
accordance with the instructions provided by ResMed. Applicable ResMed warranty
details are provided with the device at the time of original supply. Of course, as with
all electrical devices, if any irregularity becomes apparent, you should exercise caution
and have the device inspected by an Authorised ResMed Service Centre.