ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation

Page 40

Advertising
background image

36

!

CAUTION

Do not use bleach, chlorine-, alcohol- or aromatic-based solutions (including all
scented oils), moisturising or antibacterial soaps to clean the cushion, mask, air
tubing or the A

UTO

S

ET

CS2. These solutions may cause hardening and reduce

the life of the product.

Do not wash or dry the mask frame at a temperature above 80

o

C (176

o

F).

Exposure to higher temperatures may reduce the life of the product.

Do not hang the air tubing and pressure sensor tube in direct sunlight as the
tubing may harden over time and eventually crack.

P

ERIODICALLY

1. The mask and air tubing are subject to normal wear and tear. Inspect them

regularly for damage.

2. Open the power cord locking clip and remove the cord. Clean the exterior of the

flow generator with a damp cloth and mild detergent.

3. Inspect the air filter to check if it is blocked by dirt or contains holes. See full

instructions below.

!

WARNING

Beware of electric shock. Do not immerse the flow generator or power cord in
water. Always unplug the flow generator before cleaning and be sure that it is dry
before reconnecting.

!

CAUTION

Do not attempt to open the A

UTO

S

ET

CS2. There are no user serviceable parts

inside. Repairs and internal servicing should only be performed by an authorised
service agent.

A

IR

F

ILTER

Inspect the air filter every month and check if it is blocked by dirt or contains holes.

The filter should be replaced every 6 months, or more often if the flow generator is

operating in a dusty environment.

Air filter cover

eng.book Page 36 Friday, November 3, 2006 2:43 PM

Advertising