ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation

Page 37

Advertising
background image

33

T

HE

A

LARMS

E

NG

LI

SH

W

ILL

STOP

WHEN

:

the mask leak has been rectified.

N

OTE

The High Leak alarm can be turned off in the Leak Alert option in the Settings menu.

I

NSUFFICIENT

PRESSURE

SUPPORT

ALARM

T

RIGGERED

BY

:

air pressure at the mask has failed to reach an expected level for 3 breaths

the Air Circuit Learn function (page 26) has not been run after adding a new

component (eg new mask, humidifier) to the system.

I

F

AN

I

NSUFFICIENT

PRESSURE

SUPPORT

ALARM

SOUNDS

:

check that the pressure sensor tubing is not kinked

check that the pressure sensor tubing is properly connected

run LEARN CIRCUIT.

W

ILL

STOP

WHEN

:

the pressure sensor tubing is unobstructed and properly connected to the flow

generator and mask

treatment is stopped (

Start/Stop key or S

MART

S

TOP

)

the A

UTO

S

ET

CS2 is turned off.

F

LOW

BLOCKED

ALARM

T

RIGGERED

BY

:

blockage in air circuit

I

F

A

F

LOW

BLOCKED

ALARM

SOUNDS

:

1. The treatment will stop.
2. Turn power off.
3. Check whether there is a blockage in the air circuit.
4. Remove blockage.
5. Turn power back on.
6. If the Flow blocked alarm activates repeatedly, discontinue use and return to

ResMed for servicing. If the alarm does not recur, then continue to use as normal.

W

ILL

STOP

WHEN

:

blockage is removed.

eng.book Page 33 Friday, November 3, 2006 2:43 PM

Advertising