ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation
Page 37

33
T
HE
A
LARMS
E
NG
LI
SH
W
ILL
STOP
WHEN
:
•
the mask leak has been rectified.
N
OTE
The High Leak alarm can be turned off in the Leak Alert option in the Settings menu.
I
NSUFFICIENT
PRESSURE
SUPPORT
ALARM
T
RIGGERED
BY
:
•
air pressure at the mask has failed to reach an expected level for 3 breaths
•
the Air Circuit Learn function (page 26) has not been run after adding a new
component (eg new mask, humidifier) to the system.
I
F
AN
I
NSUFFICIENT
PRESSURE
SUPPORT
ALARM
SOUNDS
:
•
check that the pressure sensor tubing is not kinked
•
check that the pressure sensor tubing is properly connected
•
run LEARN CIRCUIT.
W
ILL
STOP
WHEN
:
•
the pressure sensor tubing is unobstructed and properly connected to the flow
generator and mask
•
treatment is stopped (
Start/Stop key or S
MART
S
TOP
)
•
the A
UTO
S
ET
CS2 is turned off.
F
LOW
BLOCKED
ALARM
T
RIGGERED
BY
:
•
blockage in air circuit
I
F
A
F
LOW
BLOCKED
ALARM
SOUNDS
:
1. The treatment will stop.
2. Turn power off.
3. Check whether there is a blockage in the air circuit.
4. Remove blockage.
5. Turn power back on.
6. If the Flow blocked alarm activates repeatedly, discontinue use and return to
ResMed for servicing. If the alarm does not recur, then continue to use as normal.
W
ILL
STOP
WHEN
:
•
blockage is removed.