Echnical, Pecifications – ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation
Page 47

43
T
ECHNICAL
S
PECIFICATIONS
E
NG
LI
SH
T
ECHNICAL
S
PECIFICATIONS
D
IMENSIONS
(H
X
W
X
D)
142 x 248 x 293mm (5.6" x 9.8" x 11.5")
W
EIGHT
Flow generator with cover: 3.7kg
Flow generator with H
UMIDAIRE
2
I
(empty): 4.2kg
E
NVIRONMENTAL
CONDITIONS
Operating temperature: +5
o
C to +40
o
C
Storage and transport temperature: -20
o
C to +60
o
C
Humidity — operating, storage and transportation: 10%–95% non-condensing
H
OUSING
CONSTRUCTION
Injection moulded plastic
A
IR
TUBING
1 x 2m
A
IR
FILTER
Synthetic fibre
P
ATIENT
CONNECTION
PORT
22mm taper, compatible with EN 1281-1:1997 Anaesthetic & Respiratory Equipment
- Conical Connectors
P
OWER
SUPPLY
AC Input range: 110–120 V and 220–240 V; 50–60 Hz; 60 VA
DC: 30V (via ResMed DC/DC converter)
P
ERFORMANCE
Operating pressure range: 4–20 cmH
2
O
S
OUND
PRESSURE
LEVEL
<30 dB (tested in accordance with the requirements of ISO 17510-1)
B
ATTERY
TYPE
Lithium
A
LARM
SPECIFICATIONS
All alarms comply with EN 475 for Medium priority alarms: 3 beeps (935 Hz) repeated
at intervals of 25 seconds; yellow LED flashes.
•
H
IGH
MAXIMUM
MASK
PRESSURE
ALARM
Stops air delivery at 25 cmH
2
O >700 msec.
•
H
IGH
MEAN
MASK
PRESSURE
ALARM
Stops air delivery when mean mask pressure exceeds 15 cmH
2
O (averaged over
approx 1 minute).
•
L
OW
MASK
PRESSURE
ALARM
Mask pressure is less than 3 cmH
2
O for longer than 5 seconds while the patient is
using the A
UTO
S
ET
CS2.
eng.book Page 43 Friday, November 3, 2006 2:43 PM