Echnical, Pecifications – ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation

Page 47

Advertising
background image

43

T

ECHNICAL

S

PECIFICATIONS

E

NG

LI

SH

T

ECHNICAL

S

PECIFICATIONS

D

IMENSIONS

(H

X

W

X

D)

142 x 248 x 293mm (5.6" x 9.8" x 11.5")

W

EIGHT

Flow generator with cover: 3.7kg
Flow generator with H

UMIDAIRE

2

I

(empty): 4.2kg

E

NVIRONMENTAL

CONDITIONS

Operating temperature: +5

o

C to +40

o

C

Storage and transport temperature: -20

o

C to +60

o

C

Humidity — operating, storage and transportation: 10%–95% non-condensing

H

OUSING

CONSTRUCTION

Injection moulded plastic

A

IR

TUBING

1 x 2m

A

IR

FILTER

Synthetic fibre

P

ATIENT

CONNECTION

PORT

22mm taper, compatible with EN 1281-1:1997 Anaesthetic & Respiratory Equipment

- Conical Connectors

P

OWER

SUPPLY

AC Input range: 110–120 V and 220–240 V; 50–60 Hz; 60 VA
DC: 30V (via ResMed DC/DC converter)

P

ERFORMANCE

Operating pressure range: 4–20 cmH

2

O

S

OUND

PRESSURE

LEVEL

<30 dB (tested in accordance with the requirements of ISO 17510-1)

B

ATTERY

TYPE

Lithium

A

LARM

SPECIFICATIONS

All alarms comply with EN 475 for Medium priority alarms: 3 beeps (935 Hz) repeated
at intervals of 25 seconds; yellow LED flashes.

H

IGH

MAXIMUM

MASK

PRESSURE

ALARM

Stops air delivery at 25 cmH

2

O >700 msec.

H

IGH

MEAN

MASK

PRESSURE

ALARM

Stops air delivery when mean mask pressure exceeds 15 cmH

2

O (averaged over

approx 1 minute).

L

OW

MASK

PRESSURE

ALARM

Mask pressure is less than 3 cmH

2

O for longer than 5 seconds while the patient is

using the A

UTO

S

ET

CS2.

eng.book Page 43 Friday, November 3, 2006 2:43 PM

Advertising