If oxygen is being added at the mask, the oxygen tube is attached to the mask port.
Refer to your mask manual for full instructions.
Up to 15 L/min of oxygen can be added at the mask when using an A

UTO

S

ET

CS2

system.

Figure 11: Oxygen tube attached to Mirage FFMS2 and Ultra Mirage masks

!

WARNINGS

If oxygen is used with this device, the oxygen flow must be turned off when the

device is not operating.

Explanation: When the device is not in operation, and the oxygen flow is left

on, oxygen delivered into the ventilator tubing may accumulate within the

device enclosure and create a risk of fire.

Always begin A

UTO

S

ET

CS2 therapy before the oxygen supply is turned on.

Always turn the oxygen supply off before stopping A

UTO

S

ET

CS2 therapy.

Oxygen supports combustion. Oxygen should not be used while smoking or in

the presence of an open flame.

Oxygen tube

Oxygen tube