ZOLL AED Pro Rev G Manuel d'utilisation

Page 102

Advertising
background image

A

NNEXE

A

S

PÉCIFICATIONS

A–16

www.zoll.com

9650-0350-02 Rév. G

Les performances relatives aux arythmies sont rapportées conformément à l’article, RE Kerber,
LB Becker, JD Bourland, RO Cummins, AP Hallstrom, MB Michos, G Nichol, JP Ornato,
WH Thies, RD White, BD Zuckerman, « Automated External Defibrillators for Public
Access Defibrillation: Recommendations for Specifying and Reporting Arrhythmia Analysis
Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety, » Circ J Am
Heart Assoc.
1997;95:1677-1682.

Références

Young KD, Lewis RJ. « What is confidence? Part 2: Detailed definition and determination
of confidence intervals ». Ann Emerg Med. September 1997;30:311-318.

Beyer WH. Percentage Points, F-Distribution Table. CRC Standard Mathematical Tables.
28

th

ed. Boca Raton, Fla: CRC Press; 1981:573.

Tableau A-5.

Résultats relatifs aux performances cliniques (patients enfants)

Rythmes

Taille de

l’échantillon

(sur 9 secondes)

Objectifs de

performances

Performances

observées

Limite inférieure

de l’intervalle

de confiance à
90 % unilatéral

Choquable
(49 patients)

Sensibilité

FV brute

42

>90 %

100 % (42/42)

93,1 %

TV rapide

82

>75 %

93,9 % (77/82)

87,6 %

Non choquable
(155 patients)

Spécificité

RSN

208

>99 %

100 %

(208/208)

98,6 %

FA, BS, TSV

a

,

bloc cardiaque
idioventriculaire,
ESV

a. 161 enregistrements de rythmes anormaux sur 348 concernaient des TSV (72 patients).

Les fréquences cardiaques en cas de TSV sont comprises entre 152 et 302 battements
par minute.

348

>95 %

99,4 %

(346/348)

98,2 %

Asystolie

29

>95 %

100 % (29/29)

90,2 %

Intermédiaire
(16 patients)

FV fine

0

Uniquement
rapportées

Autres TV

40

Uniquement
rapportées

90 % (36/40)

78,6 %

Advertising