Physio-Control LUCAS Manuel d'utilisation
Page 28

28
100359-50 B, ©J
OLIFE
AB 2006
LUCAS
TM
I
NSTRUCTIONS
FOR
U
SE
G
EBRUIKSAANWIJZING
I
NSTRUCTIONS
D
’
UTILISATION
WARNING – PATIENT INJURY OR DEATH
Do not leave the patient or device unattended while
LUCAS Chest Compression System is active.
2.3 C
ONNECTION
TO
AIR
SUPPLY
WARNING – INTERRUPTING AIR SUPPLY TO
LIFE SUPPORT DEVICES
If other life support equipment is disconnected from
the air supply, serious injury or death may result.
Never remove attachments to air outlets without
first checking that this will not put a patient at risk.
WARNING – EMPTY AIR SUPPLY
When the air in the air cylinder runs out, the
pressure falls and compressions delivered by LUCAS
lose force.
Immediately start manual chest compressions and
ensure that the air cylinder is exchanged. Always
change air cylinder if the pressure gauge is on red.