Chapter 6: Troubleshooting

Product Service 1-800-522-7658

26

26

PROBLEM

CAUSE

CORRECTIVE ACTION

Washer will not
drain

Clogged drain

Drain hose too high
Or hanging too low

Clean out drain.

Reroute drain hose

Filter blocked

Check filter in washer and clean as needed.

Drain valve or air
valve not operating

Check operation.

Washer overfills

Defective fill switch

Check operation.

Selection buttons
on control panel
inoperable or
controls
malfunction

Microprocessor
memory error

Perform "Factory Reset." Press and hold MENU
button and simultaneously turn on power switch.
Release MENU button and display will show the
diagnostics screen. Press MENU until Factory
Reset
is displayed. Press SELECT. Press MENU
Return to Menu is displayed. Press SELECT.

Key pad defective

Replace key pad.

Water Low Alarm

Improper water
supply

Make sure water is turned on.

Water pressure must be at least 8 psi
(55kPa) and flow must be at least 1.25 gallons
(4.7 liters)/minute.

At least .67 gallons (2.5 liters) must be available
if using pure water.

Defective fill valve

Clean screens inside the inlet fittings of the water
valve.

Check voltage to valve.

Replace valve if it does not operate with proper
voltage.

Restricted flow

Check for kinks in water hoses in base of washer
to water inlet fitting on side of tank.

Pipettes do not dry

Dry time to short

Increase dry time.

Defective air valve

Check operation in diagnostics

Check voltage to valve



Temp Fail Alarm



Temperature sensor
failure

Replace valve if it does not operate with proper
voltage.
Check connection.
Replace sensor