Chapter 2: Prerequisites

Product Service 1-800-522-7658

6

6

Compressed Air/Inert Gas Requirements


The 3/8" (9.6 mm) O.D. x 6’ (1.8m) long polyethylene tube on the rear of the
ScrubAir Pipette Washer/Dryer should be attached to a regulated source of air, or
other suitable inert gas. This tube can be shortened as required. Pressure must be
at least 20 psi (138kPa) and a minimum flow rate of 5 cfm (142 l/m). The
pressure must not exceed 100 psi (689kPa). A suitable push-to-connect or
compression fitting must be supplied by the user. An air shut off valve or quick
disconnect air fitting is recommended.
An optional 5 micron Air Filter part number 4614900 is available to help filter
incoming air if needed.