Under normal operation, your SrubAir Pipette Washer/Dryer will require little
routine maintenance to keep it functioning properly.

The following maintenance schedule is recommended. Before servicing the
ScrubAir disconnect electrical power and wear appropriate eyewear, gloves and
other safety apparel.

As needed:

Before using any cleaning or decontamination method except those recommended
by the manufacturer, users should check with the manufacturer that the proposed
method would not damage equipment.

1. Clean lid and gasket using soft cloth, sponge or chamois and a mild, non-

abrasive soap or detergent.

2. Inspect the chamber to ensure that there are no cracks or structural

damage. Call Labconco if defects exist.

3. Clean the center screen located in the bottom of the sump. Remove the

Clear Cover and Back Splash. Remove the Tube Guide Assembly from
inside the sump by lifting straight up. The screen is located in the bottom
center of the sump. Clean any debris blocking water flow through the
screen. Reassemble the washer.

Annually:

1. Check all rubber hoses and gaskets for leaks and replace any that show signs

of hardening, permanent set or deterioration.

2. Using a soft cloth, sponge, or chamois and a mild, non-abrasive soap or

detergent, clean the exterior surfaces. Liquid spray cleaners and polishes may
be used on the exterior surfaces. Do not use solvents to remove stains from
the exterior surfaces as they may damage the finish.