Appendix F: Disassembly & Reassembly Instructions

Product Service 1-800-522-7658

68

Original instructions

Transport and Reassembly

The various subassemblies are now ready to be transported and reassembled in the
opposite way the hood was disassembled. All hoods will require a minimum of five
people to lift the hood, two people on each side and one in front (for balance only). A
hydraulic lift table should be used as an aid in lifting. The 16' hood should be assembled
where placed. If in doubt about lifting the hood, always reassemble the hood where it
will be placed.

Reinstallation

Make sure the anti-rack shaft does not slide from left to right as this will be noisy and not
work effectively.

Sash Weight Reassembly

After the hood is assembled, the sash weights are easily installed before the hood is
placed against a rear wall. Please keep this in mind and install the sash weights now.

Baffles Installation

Install the upper baffle, middle baffle, and lower baffle. Be sure the baffles are resting in
the proper baffle mount supports.

Hood Replacement and Service Installation

The hood is now ready for placement and the HVAC, plumbing and electrical service can
be installed.