Chapter 2: Prerequisites
Product Service 1-800-522-7658
9
Original instructions
Proper blower selection for vertical and horizontal floor-mounted hoods can be
determined from these exhaust requirements and the total system static pressure
loss. Contact Labconco Customer Service for assistance in sizing a blower
system.
Projected Airflows and Static Pressure
Hood Type
Face
Velocity (fpm) Airflow Volumetric Rate (CFM) @
Vertical
Top Sash
Top Sash
Static Pressure (inches of water)
Floor
Full open
open 62.5%
4'
5'
6'
8'
Mount
at 64"
at 54"
Hood
Hood
Hood
Hood
Walk-In
100
160
725, 0.27"
955, 0.34" 1180, 0.46" 1640, 0.31"
XL
80
128
580, 0.17"
765, 0.22"
945, 0.29" 1310, 0.20"
W/Bypass
62.5
100
450, 0.11"
595, 0.13"
735, 0.18" 1025, 0.12"
Block
60
96
435, 0.10"
575, 0.12"
710, 0.17"
985, 0.11"
50
80
365, 0.07"
480, 0.09"
590, 0.11"
820, 0.08"
40
64
290, 0.05"
380, 0.06"
470, 0.08"
655, 0.05"
38
60
270, 0.04"
360, 0.05"
440, 0.07"
615, 0.04"
Note: Bottom Sash is closed and
only opened for loading.
Projected Airflows and Static Pressure
Doors Open
2 Doors
1 Door
1 Door
1 Door
Hood Type
Face
Velocity (fpm) Airflow Volumetric Rate (CFM) @
Horizontal Horizontal
Static Pressure (inches of water)
Floor
Sash at
8'
10'
12'
16'
Mount
Full Open
Hood
Hood
Hood
Hood
Walk-In
XL
100
2150, 0.53" 1450, 0.22" 1765, 0.30" 2400, 0.18"
W/Bypass
80
1720, 0.34" 1160, 0.14" 1410, 0.19" 1920, 0.12"
Block
2 Doors
1 Door
1 Door
1 Door
Open
Open
Open
Open