Chapter 2: Prerequisites

Product Service 1-800-522-7658

9

Original instructions

Proper blower selection for vertical and horizontal floor-mounted hoods can be
determined from these exhaust requirements and the total system static pressure
loss. Contact Labconco Customer Service for assistance in sizing a blower
system.

Projected Airflows and Static Pressure

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Vertical

Top Sash

Top Sash

Static Pressure (inches of water)

Floor

Full open

open 62.5%

4'

5'

6'

8'

Mount

at 64"

at 54"

Hood

Hood

Hood

Hood

Walk-In

100

160

725, 0.27"

955, 0.34" 1180, 0.46" 1640, 0.31"

XL

80

128

580, 0.17"

765, 0.22"

945, 0.29" 1310, 0.20"

W/Bypass

62.5

100

450, 0.11"

595, 0.13"

735, 0.18" 1025, 0.12"

Block

60

96

435, 0.10"

575, 0.12"

710, 0.17"

985, 0.11"

50

80

365, 0.07"

480, 0.09"

590, 0.11"

820, 0.08"

40

64

290, 0.05"

380, 0.06"

470, 0.08"

655, 0.05"

38

60

270, 0.04"

360, 0.05"

440, 0.07"

615, 0.04"

Note: Bottom Sash is closed and

only opened for loading.

Projected Airflows and Static Pressure

Doors Open

2 Doors

1 Door

1 Door

1 Door

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Horizontal Horizontal

Static Pressure (inches of water)

Floor

Sash at

8'

10'

12'

16'

Mount

Full Open

Hood

Hood

Hood

Hood

Walk-In

XL

100

2150, 0.53" 1450, 0.22" 1765, 0.30" 2400, 0.18"

W/Bypass

80

1720, 0.34" 1160, 0.14" 1410, 0.19" 1920, 0.12"

Block

2 Doors

1 Door

1 Door

1 Door

Open

Open

Open

Open