PPENDIX

F

D

ISASSEMBLY

AND

R

EASSEMBLY

I

NSTRUCTIONS

Required Tools

The following tools will be required for disassembly and reassembly:

Cordless Drills and Phillips bits.

1/4", 5/16", 3/8", 7/16" Sockets

Small Screwdriver

Phillips Screwdriver

Right Angle Screwdriver

1/4", 5/16" Box Ratchet

1/4" Extension

5/16", 3/8", 7/16" Combination Wrench 1/4" Ratchet

Small Right Angle Needle Nose Pliers Nut Driver Set

Decorative Panel Removal

1. Remove side panels by removing #10-24 screws and lifting away.

2. Remove front panels by lifting up and away.


Light Removal and Blower Wiring

1. On the top of the hood, disconnect the Molex connector and remove the green ground

wire. All the removable wires are part of the wiring harness to the corner post
switches. Be sure to label all wires and refer to the wiring diagrams for proper
reassembly.


Header Removal

1. The header is fastened to the corner posts by four #12 screws. Simply reach behind

the corner posts from the side and remove the screws while another person supports
the header.