Baffle installation for the floor-mounted hoods – Labconco Protector XL Floor-Mounted (Walk-In) Laboratory Fume Hoods 1128 Series Manuel d'utilisation

Page 25

Advertising
background image

Chapter 3: Getting Started

Product Service 1-800-522-7658

21

Original instructions

















Figure 3-6

Corner Post Installation for the Floor-

Mounted Hoods

Both left and right corner posts have been shipped uninstalled. The one-piece
corner posts are to be installed once the upper and lower cabinet assemblies have
been properly positioned.

The edges on the corner posts fit directly onto the side frames. The front inner
edge of both corner posts are held in place by stainless steel machine screws. The
outer back edge of both corner posts are held in place with steel self-drilling
screws. The screws are included in the hood manual packet. See Figure 3-2 and
Figure 3-3.

Baffle Installation for the Floor-Mounted

Hoods

The baffles are critical to the proper airflow and performance of the hood. Be
sure the baffles are resting in the proper baffle mount supports. See Figure 3-7 for
various sizes of baffles and their orientation.

Sliding Sash Door

Lower Sash Track
Threshold

Center Support
for Header
Sash Track

Upper Header Sash Track

End Support for
Header Sash Track