SYMBOLS

56

Symbols

SYMBOL

WHAT IT MEANS

WHAT TO DO

Remove the test strip from the meter. If white patches or streaks
appear on the confirmation dot, you have not applied enough blood
to obtain an accurate test. You must start over with a new test strip.
When the “Insert Test Strip” symbol appears on the display, insert
the test strip. Push it firmly and more slowly into the test strip holder
until it stops.

Note:

It is possible to push the test strip into the test strip holder

too fast. If this happens, the meter displays the “Remove Test
Strip” symbol.

Remove
Test Strip

Firmly push the test strip into the test strip holder, with the
tapered-end first and the confirmation dot facing down, until it
stops. If you fail to completely insert the test strip, the test may
start. However, you may receive an inaccurately low result.

Insert Test
Strip

The following symbols appear on the meter display to help guide you through the testing
process. If they occur during testing, they may indicate problems.