NOTE: If the actions listed below do not resolve the error, call the OneTouch

®

Customer Care Line at 1 800 663-5521.

ERROR

POTENTIAL PROBLEM

WHAT TO DO

1.

Repeat the test with a new test strip and:

Check the test strip expiration date. Do not use
test strips beyond the expiration or discard
date.

The meter detected
a problem with the
test strip.

Professional Use:

Test the sample with an alternate method.
While awaiting results:
Compare the colour of the test strip confirmation dot
to the Colour Chart on the test strip vial label. If the
glucose level is very low (2.8 mmol/L) or very high
(19.4 mmol/L), take appropriate precautions
immediately.
Before retesting verify that the sample hematocrit is
within range.

If these actions do not resolve the error, call the
OneTouch

®

Customer Care Line at 1 800 663–5521.

The meter detected
a problem with the
colour of the test
strip.

Er 1

continued

50

TROUBLESHOOTING

DCO_05090602A_OTSS_OB_CA_ENG 12/3/03 1:27 PM Page 50