1.

Press the blue power button to turn on
the meter. The display check appears
briefly. Carefully check that all symbols
appear. Make sure a test strip is not in
the meter when you turn it on.

2.

Compare the code number on the meter
display with the code number on the
test strip vial. If they do not match,
press the green

C

button until the

codes match.

3.

Remove a test strip from the vial and
replace the vial cap. Close it tightly.

Check the expiration date and discard
date. Do not use the test strips after
these dates.

Compare the colour of the confirmation
dot on the test strip with the “Unused”
colour dot shown on the Colour Chart
on the test strip vial label. If the colour
of the confirmation dot on the test strip
is darker than that shown on the Colour
Chart, do not use the test strip. The
result may be inaccurate.