background image

Chapter 4: Performance Data and Safety Precautions

Product Service 1-800-522-7658

27

Original instructions

Speed

(RPM)

Optimum Airflow (CFM) and Static Pressure (inches of water)

1500 MAX

900 CFM @ 1.9" to 1025 CFM @ 1.8"

1400

900 CFM @ 1.8" to 1180 CFM @ 1.7"

1250

850 CFM @ 1.37" to 1325 CFM @ 0.8"

1000

675 CFM @ 0.9" to 1100 CFM @ 0.6"

750

450 CFM @ 0.45" to 825 CFM @ 0.33"

500

300 CFM @ 0.20" to 550 CFM @ 0.12"

300

200 CFM @ 0.08" to 400 CFM @ 0.10"

Practical range is 200 CFM, 0.08" to 1180 CFM, 1.7"

0

0.25

0.5

0.75

1

1.25

1.5

1.75

2

2.25

150 250 350 450 550 650 750 850 950 1050 1150 1250 1350 1450 1550

St

ati

c

Pressure

(in.

w

.g.)

Airflow Rate (CFM)

Models 71834-12, -13, -22, -23

120- 240 Volts 10.0 Amps 1 HP

Blower Performance Curve with 12" PVC inlet

500

750

1000

RPM

1250 RPM

1500

Optimum Blower

Performance

Note: Ambient Temperatures

Inside Motor Housing Greater

than 125 °F May Reduce

Maximum Blower Performance