Chapter 4: Performance Data and Safety Precautions

Product Service 1-800-522-7658

25

Original instructions

Speed

(RPM)

Optimum Airflow (CFM) and Static Pressure (inches of water)

1400 MAX

960 CFM @ 1.7" to 1100 CFM @ 1.7"

1300

900 CFM @ 1.5" to 1150 CFM @ 1.4"

1250

850 CFM @ 1.3" to 1400 CFM @ 0.87"

1125

780 CFM @ 1.1" to 1310 CFM @ 0.75"

1000

700 CFM @ 0.87" to 1180 CFM @ 0.62"

750

450 CFM @ 0.4" to 860 CFM @ 0.34"

500

300 CFM @ 0.20" to 575 CFM @ 0.12"

300

200 CFM @ 0.08" to 450 CFM @ 0.10"

Practical range is 200 CFM, 0.08" to 1400 CFM, 0.87" or 1150 CFM, 1.4"

0

0.25

0.5

0.75

1

1.25

1.5

1.75

2

2.25

150 250 350 450 550 650 750 850 950 1050 1150 1250 1350 1450 1550

St

ati

c

Pressure

(in.

w

.g.)

Airflow Rate (CFM)

Models 71818-12, -13, -22, -23

120- 240 Volts 10.0 Amps 1 HP

Blower Performance Curve with 12" FRP inlet

500

750

1000

1250 RPM

1400

1500

Optimum Blower

Performance

Note: Ambient Temperatures

Inside Motor Housing Greater

than 125 °F May Reduce

Maximum Blower Performance

1300

RPM

See Note