Labconco Intelli-Sense Multi-Speed Blowers 7183413 Manuel d'utilisation

Page 27

Advertising
background image

Chapter 4: Performance Data and Safety Precautions

Product Service 1-800-522-7658

23

Original instructions

Speed

(RPM)

Optimum Airflow (CFM) and Static Pressure (inches of water)

1125 MAX

1200 CFM @ 1.6" to 1350 CFM @ 1.5"

1000

1100 CFM @ 1.3" to 1450 CFM @ 1.2"

750

800 CFM @ 0.8" to 1150 CFM @ 0.7"

500

550 CFM @ 0.33" to 725 CFM @ 0.3"

300

200 CFM @ 0.08" to 450 CFM @ 0.1"

Practical range is 200 CFM, 0.08" to 1350 CFM, 1.5" or 1450 CFM @ 1.2"

0

0.25

0.5

0.75

1

1.25

1.5

1.75

2

2.25

150 250 350 450 550 650 750 850 950 1050 1150 1250 1350 1450 1550

St

ati

c

Pressure

(in.

w

.g.)

Airflow Rate (CFM)

500

750

1000

RPM

Note: Ambient Temperatures

Inside Motor Housing Greater

than 125 °F May Reduce

Maximum Blower Performance

Models 70611-12, -13, -22, -23

120- 240 Volts 10.0 Amps 1 HP

Blower Performance Curve

with 12" Coated Steel inlet

1250 RPM

See

Note

Optimum

Blower

Performance

Advertising