Chapter 3: Getting Started

Product Service 1-800-522-7658

14

Original instructions

5.

Run the MSB Control Box Power Cable with the IEC inlet from the control

box to the junction box on the fume hood and connect to the fume hood
light circuit per wiring detail and wirings diagram in Appendix C.

6.

If two or three position blower setting is desired, remove the blower switch

and blower switch label from the fume hood. For simple ON/OFF, skip
this step.

7.

Remove the wires to the blower switch. Important: Prevent the existing

wires to the blower switch from causing interference with the sash or
remove them.

8.

Install the new blower switch label, blower switch, and blower switch

retainer. If needed, blower switch cutout is 0.83" x 1.45" (2.1 cm x 3.7 cm).

9.

Locate the MSB Control Cable and connect the wires from the blower

switch to the MSB Control Box connector per wiring detail and wiring
diagrams in Appendix C and reference Appendix E.

10.

All electrical connections to the roof mounted blower, the MSB Control

Box, and switch wiring should be intact.

11.

Restore power to the roof mounted blower and the fume hood lighting

circuit.