Optimum blower performance – Labconco Intelli-Sense Multi-Speed Blowers 7183413 Manuel d'utilisation

Page 28

Advertising
background image

Chapter 4: Performance Data and Safety Precautions

Product Service 1-800-522-7658

24

Original instructions

Speed

(RPM)

Optimum Airflow (CFM) and Static Pressure (inches of water)

1800 MAX

750 CFM @ 1.7" to 1025 CFM @ 1.55"

1500

600 CFM @ 1.2" to 1050 CFM @ 0.5"

1250

500 CFM @ 0.8" to 850 CFM @ 0.37"

1000

400 CFM @ 0.6" to 700 CFM @ 0.25"

750

300 CFM @ 0.3" to 475 CFM @ 0.12"

500

200 CFM @ 0.12" to 325 CFM @ 0.06"

300

175 CFM @ 0.06" to 275 CFM @ 0.04"

Practical range is 175 CFM, 0.06" to 1025 CFM, 1.5"

0

0.25

0.5

0.75

1

1.25

1.5

1.75

2

150

250

350

450

550

650

750

850

950 1050 1150 1250 1350

St

ati

c

Pressure

(in.

w

.g.)

Airflow Rate (CFM)

Models 71818-10, -11, -20, -21

120- 240 Volts 10.0 Amps 1 HP

Blower Performance Curve with 10" FRP inlet

500

750

1000

Optimum

Blower

Performance

Note: Ambient Temperatures

Inside Motor Housing Greater

than 125 °F May Reduce

Maximum Blower Performance

1250

1500

1750

1800

SEE

NOTE

Advertising