Optimum blower performance – Labconco Intelli-Sense Multi-Speed Blowers 7183413 Manuel d'utilisation

Page 26

Advertising
background image

Chapter 4: Performance Data and Safety Precautions

Product Service 1-800-522-7658

22

Original instructions

Speed

(RPM)

Optimum Airflow (CFM) and Static Pressure (inches of water)

1800 MAX

750 CFM @ 2.2" to 1050 CFM @ 1.9"

1500

650 CFM @ 1.6" to 1250 CFM @ 0.7"

1250

550 CFM @ 1.1" to 1050 CFM @ 0.5"

1000

400 CFM @ 0.7" to 850 CFM @ 0.33"

750

275 CFM @ 0.37" to 650 CFM @ 0.20"

500

200 CFM @ 0.20" to 450 CFM @ 0.10"

300

175 CFM @ 0.06" to 350 CFM @ 0.08"

Practical range is 175 CFM, 0.06" to 1050 CFM, 1.9" or 1200 CFM, 1.0"

0

0.25

0.5

0.75

1

1.25

1.5

1.75

2

2.25

2.5

150 250 350 450 550 650 750 850 950 1050 1150 1250 1350 1450

Stat

ic

Pressur

e

(in.

w

.g.

)

Airflow Rate (CFM)

Models 70611-10, -11, -20, -21

120- 240 Volts 10.0 Amps 1 HP

Blower Performance Curve with 10" Coated Steel inlet

500

750

1000

Optimum

Blower

Performance

Note: Ambient Temperatures

Inside Motor Housing Greater

than 125 °F May Reduce

Maximum Blower Performance

1250

1500

1750

SEE

NOTE

Advertising