Chapter 3: Getting Started

Product Service 1-800-522-7658

11

Original instructions

haubans doivent être suffisamment importants pour soutenir la
cheminée d'échappement contre la vitesse des vents élevés.

HINT: An exhaust damper either before or after the MSB
Blower is very useful for fine tuning the airflow.

For Coated Steel Blowers


10" Coated Steel Blowers include a 10" by 5-1/2" rectangular outlet. A
rectangular to round Transition Adaptor, Labconco part number 4722401, adapts
the outlet connection to accept 10-inch diameter PVC ductwork Transition
Adaptor. Labconco part number 4722400 adapts the outlet on these blowers to
accept 8-inch diameter PVC ductwork.

12" Coated Steel Blowers include a 13-1/2" by 7" rectangular outlet. Transition
Adaptor, Labconco part number 7003400, adapts the outlet of these blowers to
accept 12-inch diameter PVC ductwork.

For Fiberglass Blowers


10" Fiberglass Blowers feature a 10-3/4" ID outlet connection. Ten-inch diameter
ductwork will slip into this connection and should be held by sheet metal screws
through the housing. Silicone sealant should be used to seal any air leaks between
the duct and blower outlet connection.

12" Fiberglass Blowers feature a 12-3/4" ID outlet connection. Twelve-inch
diameter ductwork will slip into this connection and should be held by sheet metal
screws through the housing. Silicone sealant should be used to seal any air leaks
between the duct and blower outlet connection.

For PVC Blowers


10" PVC Blowers feature a 10-3/4" ID outlet connection. Ten-inch diameter
ductwork will slip into this connection and should be held by sheet metal screws
through the housing. Silicone sealant should be used to seal any air leaks between
the duct and blower outlet connection.

12" PVC Blowers feature a 12-3/4" ID outlet connection. Twelve-inch diameter
ductwork will slip into this connection and should be held by sheet metal screws
through the housing. Silicone sealant should be used to seal any air leaks between
the duct and blower outlet connection.

CS

FRP

PVC