Appendix c: waterpro ps system specifications – Labconco WaterPr PS Systems 90007 Series Manuel d'utilisation

Page 45

Advertising
background image

Appendix C: WaterPro PS System Specifications

41

Original Instructions


HPLC/UF Hybrid models
915 Grains as CaCO

3

Deionization:

High Purity Polishing grade mixed bed resin,

which will deliver 16 to 18.2 Megohm.cm Type I
water.

Ultrafiltration:
(membrane included on UF models)

Polysulfone membrane in a spirally wound

configuration.

Final Filtration (Optional):

Self-venting 0.2 micron hollow fiber filter

Electrical Specifications:

115V, 60 Hz, 5.0 Amps or 230V, 50 Hz, 2.5 Amps

Single Phase

Relative Humidity:

Less than 80%


Environmental Conditions:

The WaterPro PS is designed to operate safely under the following conditions:

Indoor use

Altitude up to 2,000M (6,562 Ft.)

Ambient temperatures 5°C to 40°C (41°F to 104°F)

Maximum relative humidity 80% for temperatures up to 31°C (88°F) decreasing linearly to
50% relative humidity at 40°C (104°F)

Main supply voltage fluctuations not to exceed ±10% of the nominal voltage

Transient over-voltages according to installation category II (over-voltage categories per
IEC 1010)

Pollution degrees 2 (Normally only non-conductive foreign matter, solid, liquid, or gaseous
(ionized gases), that may produce a reduction of dielectric strength or surface resistivity
occurs. Occasionally, however, a temporary conductivity caused by condensation must be
expected, in accordance with IEC 664)