Chapter 6: Troubleshooting

31

Original Instructions

PROBLEM

CAUSES

CORRECTIVE

ACTION

Reduced cartridge
life

Cartridges are past expiration

date

Check the manufactured date on

the cartridge package. The shelf
life of unopened cartridges is one
year from the date of manufacture.
Replace the expired cartridges
with new ones.

Feed-water quality has changed

If the source is a reverse osmosis

system, ensure that that it is
working properly.

If the source is a central water

supply, verify that it is still
working properly. Check the
average tap water supply versus
the deionization capacity chart on
page 43.

Odor from
polished water

Bacterial growth on the filters

Remove old filters. Sanitize

system. Install new filters.

Old DI filters. DI resins breaking

down

Remove old filters. Sanitize

system. Install new filters.