Refer to the following if your WaterPro PS System fails to operate properly. If the suggested
corrective actions do not solve your problem, contact Labconco for additional assistance.

PROBLEM

CAUSES

CORRECTIVE

ACTION

Unit inoperative

Unit not plugged into outlet

Plug the Polishing Station into
appropriate electrical service.

Circuit breaker tripped/fuse

blown

Reset circuit breaker/replace fuse.

Power switch is in the off

position

Press power button to switch

system on.

Defective circuit board

Replace circuit board.

Inlet valve closed do to moisture

in drip pan.

Check system for leaks. Dry drip

pan.

Recirculation
pump operates but
no water is
dispensed

Customer supplied feed valve is

closed

Open the valve.

Feed-water line is restricted, or

no supply

Inspect and adjust feedwater line

as required.

Final filter is clogged

Replace final filter.