Chapter 5: Maintaining Your WaterPro PS System

26

Original Instructions


UV Lamp Replacement
(HPLC, Life Science and HPLC/UF, Hybrid Models)


Avoid direct exposure of eyes and skin to ultraviolet light.

Evitez toute exposition directe des yeux et la peau à la lumière ultraviolette.



The UV lamp should be replaced annually to maintain lamp intensity. To replace lamp, use the
following procedure (Refer to Figure 6).

(1) Unplug the WaterPro PS system and remove the front cover via the two screws on the front of

the cover.

(2) Locate the ultraviolet reactors by referring to Item 14 under Component Identification located

on Page 12 of this manual.

(3) Disconnect the two white leads and pull the bulb from the reactor.
(4) Insert new bulb Catalog number 9109200 into the reactor and reconnect the leads.

Figure 6

!